MIL-HDBK-1839A APPENDIX B
CATEGORY I OPERATIONAL/SYSTEM EQUIPMENT |
CATEGORY II TEST/MEASUREMENT & DIAGNOSTIC EQUIPMENT |
CATEGORY III CALIBRATION EQUIPMENT/STANDARDS |
||||||||
TOC No. |
Description of Item |
Operation Range Spec Value |
Operating Tolerance |
Interval |
Description of Item |
Specific Range/ Value |
Specific Tolerance |
Description of Item |
Specific Range/ Value |
Specific Tolerance |
1.1 |
Continued |
|||||||||
AC Voltage Gain Freq. @ 1 kHz (Vac) |
Meter Calibrator PN 5700A NAF (Part of PATEC) |
33K8-4-567-1 (Air Force Procedure) |
||||||||
.03 |
0.4 % |
200 mV |
0.1 % |
|||||||
0.3 |
0.4 % |
2 V |
0.1 % |
|||||||
1.0 |
0.4 % |
2 V |
0.1 % |
|||||||
2.0 |
0.4 % |
2 V |
0.1 % |
|||||||
3.0 |
0.4 % |
20 V |
0.1 % |
|||||||
30.0 |
0.4 % |
200 V |
0.1 % |
|||||||
300.0 |
0.4 % |
200 V |
0.1 % |
|||||||
30.0 |
0.4 % |
20 V |
0.1 % |
|||||||
3.0 |
1.0 % |
2 V |
0.25 % |
|||||||
0.3 |
1.0 % |
2 V |
0.25 % |
|||||||
0.03 |
1.0 % |
200 mV |
0.25 % |
|||||||
AC Voltage Frequency Response Test (Vac) |
Meter Calibrator PN: 5700A NAF (Part of PATEC) |
33K8-4-567-1 (Air Force Procedure) |
28
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business