MIL-HDBK-1839A APPENDIX B
CATEGORY I OPERATIONAL/SYSTEM EQUIPMENT |
CATEGORY II TEST/MEASUREMENT & DIAGNOSTIC EQUIPMENT |
CATEGORY III CALIBRATION EQUIPMENT/STANDARDS |
||||||||
TOC No. |
Description of Item |
Operation Range Spec Value |
Operating Tolerance |
Interval |
Description of Item |
Specific Range/ Value |
Specific Tolerance |
Description of Item |
Specific Range/ Value |
Specific Tolerance |
1.3 |
Digitizing O'Scope 250 MHz E1428A Ch 2 |
|||||||||
Channel 2 DC coupled |
50 ohms |
0.5 ohms |
30 ohms |
+ 0.125 ohms |
||||||
VOLTAGE MEASUREMENT UNCERTAINTY |
Meter Calibrator 5700A NAF (Part of PATEC) |
33K8-4-567-1 (Air Force Procedure) |
||||||||
Ch 1 & Ch 2 Range |
DCV |
|||||||||
Channel 1 with 15V offset |
40 Vdc |
0.4 % |
30 Vdc |
+ 0.1% |
||||||
Input Impedance 50 ohms |
15 Vdc |
0.4 % |
15 Vdc |
+ 0.1% |
||||||
5 Vdc |
0.4 % |
5 Vdc |
+ 0.1% |
|||||||
Offset 600 mV |
1.2 Vdc |
0.4 % |
1.2 Vdc |
+ 0.1% |
||||||
600 mVdc |
0.4 % |
600 mVdc |
+ 0.1% |
|||||||
200 mVdc |
0.4 % |
200 mVdc |
+ 0.1% |
|||||||
60 mVdc |
0.4 % |
60 mVdc |
+ 0.1% |
|||||||
Offset 30 mV |
30 mVdc |
0.4 % |
30 mVdc |
+ 0.1% |
||||||
10 mVdc |
0.4 % |
10 mVdc |
+ 0.1% |
|||||||
Offset 21 mV |
42 mVdc |
0.4 % |
42 mVdc |
+ 0.1% |
30
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business